Wednesday, June 24, 2015

Research: Development of an antimicrobial urinary catheter to inhibit urinary catheter encrustation

Rachael P. C. Jordan1*, Sladjana Malic1,2, Mark G. Waters1,3, David J. Stickler4 and David W. Williams1

*Correspondence: Rachael P. C. Jordan JordanRP@cardiff.ac.uk

1. School of Dentistry, College of Biomedical and Life Sciences, Cardiff University, Heath Park, Cardiff, CF14 4XY, Wales, UK.

[+] Author Affiliations

DOI : http://dx.doi.org/10.7243/2052-6180-3-1

© 2015 Jordan et al; licensee Herbert Publications Ltd.

This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/3.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Abstract

Background: Encrustation of urinary catheters is a frequent problem in patients with long-term indwelling catheters colonised with urease-positive bacteria such as Proteus mirabilis. Catheter blockage may follow catheter encrustation, potentially leading to systemic infection. Prevention of encrustation is difficult and avoidance of recurrence often unsuccessful. One possible preventative strategy is to use a catheter with an antimicrobial surface and development and assessment of such a surface was the aim of this research.

Methods: Initial experiments assessed the antimicrobial activity of silicone impregnated with plant-derived antimicrobials and triclosan using agar diffusion. The longevity of activity of each antimicrobial silicone was examined over a period of 11 weeks following soaking individual pieces of antimicrobial silicone in an artificial urine solution before using agar diffusion to test remaining antimicrobial activity. Live/Dead staining of bacteria colonising the surface of each antimicrobial silicone was employed to determine the bactericidal properties of each antimicrobial silicone. Selected antimicrobial silicones were subsequently evaluated for their ability to prevent catheter encrustation in an in vitro bladder model.

Results: Results showed that antimicrobial activity was obtained using 1% triclosan-impregnated silicone and that this antimicrobial activity was long-lasting (up to 11 weeks). Use of a dip coat silicone formulation, containing 0.2% triclosan, proved effective at delaying catheter encrustation with P. mirabilis metabolites in vitro. In 8 out of 13 independent experiments using dip-coated catheters, no catheter blockage occurred over 7 days, whilst all control catheters blocked during this period. Only on one occasion was delayed encrustation not evident with the treated catheters.

Conclusions: In summary, a dip-coat silicone containing triclosan proved effective in preventing in vitro catheter encrustation caused by P. mirabilis infection. Further studies with triclosan silicone dip coat formulation are warranted, including those that investigate potential host cell toxicity and long-term benefits following its application to indwelling urinary catheters in clinical settings.

KeywordsProteus mirabilis, catheter associated urinary tract infection, triclosan, eugenol, terpinen-4-ol

Read full article here:  www.hoajonline.com/microbiology/2052-6180/3/1

0 comments: